700B Headstage Calibration

點(diǎn)擊次數(shù):1134 發(fā)布時間:2015-1-28

700B Headstage Calibration

The steps below provide a quick check of the calibration of the MultiClamp 700B.  It is assumed that appropriate shielding (as described in “Test the Noise", above) will be used during these tests.

1.  Connect an oscilloscope to the front panel PRIMARY or SCOPE Output BNC.

2.  Synchronize the oscilloscope by connecting to the rear panel SYNC OUTPUT BNC. Press the “Reset to Program Defaults" button in the MultiClamp 700B Commander to standardize the MultiClamp 700B.


50 G Range

1.  Press the “Options" button, choose the Gains tab, and select the 50 G feedback resistor in the Voltage Clamp pane.  Return to the main MultiClamp 700B Commander window.

2.  Plug the PATCH connector of the PATCH-1U model cell into the CV-7 headstage.

3.  Set Seal Test amplitude to 100 mV and frequency to 50 Hz, then check the box to make it active.

4.  Press Auto Cp Fast to remove the bulk of electrode capacitance transient.

5.  The resulting waveform should be square, except for an initial overshoot (possibly twice the size of the steady-state response) that settles to the baseline in about 1 to 2 ms.  The rise time to the peak of the overshoot should be about 50 μs.  The steady-state amplitude following the transient should be ~500 mVp-p (±50 mV).


5 G Range

1.  Change the feedback resistor to 5 G in the Options / Gains menu.

2.  Press Auto Cp Fast.

3.  The step response should be ~50 mVp-p (±5 mV).


500 M Range

1.  Press the “Reset to Program Defaults" button.  (By default, the 500 M

range is selected.)

2.  Check Seal Test and set the amplitude to 25 mV.

3.  Plug the CELL connector of the PATCH-1U model cell into the CV-7 headstage.

4.  Press the Auto Whole Cell button.

5.  Press Auto Cp Fast button.

6.  The step response should be ~25 mVp-p. 


50 M Range

1.  Change the feedback resistor to 50 M.

2.  Increase Output Gain to 10.

3.  Press Auto Whole Cell and Auto Cp Fast.

4.  The step response should be ~25 mVp-p.  



[ 打印 ] [ 返回頂部 ] [ 關(guān)閉

| 商鋪首頁 | 公司檔案 | 產(chǎn)品展示 | 供應(yīng)信息 | 公司動態(tài) | 詢價留言 | 聯(lián)系我們 | 會員管理 |
化工儀器網(wǎng) 設(shè)計制作,未經(jīng)允許翻錄必究.Copyright(C) http://www.9av99.com, All rights reserved.
以上信息由企業(yè)自行提供,信息內(nèi)容的真實(shí)性、準(zhǔn)確性和合法性由相關(guān)企業(yè)負(fù)責(zé),化工儀器網(wǎng)對此不承擔(dān)任何保證責(zé)任。
溫馨提示:為規(guī)避購買風(fēng)險,建議您在購買產(chǎn)品前務(wù)必確認(rèn)供應(yīng)商資質(zhì)及產(chǎn)品質(zhì)量。
二維碼

掃一掃訪問手機(jī)站
吕梁市| 周宁县| 板桥市| 湘潭县| 广东省| 宾阳县| 徐汇区| 城口县| 平顶山市| 平定县| 巫溪县| 佛学| 镇平县| 汉中市| 陈巴尔虎旗| 屏边| 怀来县| 上林县| 宜宾县| 泸州市| 泾川县| 玛沁县| 黔西县| 前郭尔| 望谟县| 黎川县| 来凤县| 安国市| 和平县| 利辛县| 巨鹿县| 蒙山县| 额尔古纳市| 天气| 蓝田县| 烟台市| 德江县| 辽阳市| 建始县| 延长县| 贵阳市|