700B Headstage Calibration

點(diǎn)擊次數(shù):1134 發(fā)布時間:2015-1-28

700B Headstage Calibration

The steps below provide a quick check of the calibration of the MultiClamp 700B.  It is assumed that appropriate shielding (as described in “Test the Noise", above) will be used during these tests.

1.  Connect an oscilloscope to the front panel PRIMARY or SCOPE Output BNC.

2.  Synchronize the oscilloscope by connecting to the rear panel SYNC OUTPUT BNC. Press the “Reset to Program Defaults" button in the MultiClamp 700B Commander to standardize the MultiClamp 700B.


50 G Range

1.  Press the “Options" button, choose the Gains tab, and select the 50 G feedback resistor in the Voltage Clamp pane.  Return to the main MultiClamp 700B Commander window.

2.  Plug the PATCH connector of the PATCH-1U model cell into the CV-7 headstage.

3.  Set Seal Test amplitude to 100 mV and frequency to 50 Hz, then check the box to make it active.

4.  Press Auto Cp Fast to remove the bulk of electrode capacitance transient.

5.  The resulting waveform should be square, except for an initial overshoot (possibly twice the size of the steady-state response) that settles to the baseline in about 1 to 2 ms.  The rise time to the peak of the overshoot should be about 50 μs.  The steady-state amplitude following the transient should be ~500 mVp-p (±50 mV).


5 G Range

1.  Change the feedback resistor to 5 G in the Options / Gains menu.

2.  Press Auto Cp Fast.

3.  The step response should be ~50 mVp-p (±5 mV).


500 M Range

1.  Press the “Reset to Program Defaults" button.  (By default, the 500 M

range is selected.)

2.  Check Seal Test and set the amplitude to 25 mV.

3.  Plug the CELL connector of the PATCH-1U model cell into the CV-7 headstage.

4.  Press the Auto Whole Cell button.

5.  Press Auto Cp Fast button.

6.  The step response should be ~25 mVp-p. 


50 M Range

1.  Change the feedback resistor to 50 M.

2.  Increase Output Gain to 10.

3.  Press Auto Whole Cell and Auto Cp Fast.

4.  The step response should be ~25 mVp-p.  



[ 打印 ] [ 返回頂部 ] [ 關(guān)閉

| 商鋪首頁 | 公司檔案 | 產(chǎn)品展示 | 供應(yīng)信息 | 公司動態(tài) | 詢價留言 | 聯(lián)系我們 | 會員管理 |
化工儀器網(wǎng) 設(shè)計制作,未經(jīng)允許翻錄必究.Copyright(C) http://www.9av99.com, All rights reserved.
以上信息由企業(yè)自行提供,信息內(nèi)容的真實(shí)性、準(zhǔn)確性和合法性由相關(guān)企業(yè)負(fù)責(zé),化工儀器網(wǎng)對此不承擔(dān)任何保證責(zé)任。
溫馨提示:為規(guī)避購買風(fēng)險,建議您在購買產(chǎn)品前務(wù)必確認(rèn)供應(yīng)商資質(zhì)及產(chǎn)品質(zhì)量。
二維碼

掃一掃訪問手機(jī)站
合水县| 安达市| 疏附县| 古丈县| 乾安县| 东乌珠穆沁旗| 镇平县| 浮山县| 房山区| 禄丰县| 泉州市| 宁河县| 雅江县| 滨州市| 甘德县| 上犹县| 习水县| 乌拉特前旗| 天镇县| 沂南县| 璧山县| 彩票| 买车| 佛教| 浮山县| 体育| 荃湾区| 庆城县| 天门市| 揭阳市| 宁陵县| 寿宁县| 铜鼓县| 吴江市| 上杭县| 武义县| 阜平县| 通山县| 西和县| 含山县| 彭阳县|