BR-UV-PV 步入式光伏組件紫外預(yù)處理試驗室
- 公司名稱 上海泊??茖W(xué)儀器有限公司
- 品牌
- 型號 BR-UV-PV
- 產(chǎn)地 上海泊睿
- 廠商性質(zhì) 生產(chǎn)廠家
- 更新時間 2018/5/14 9:00:00
- 訪問次數(shù) 3152
聯(lián)系我們時請說明是化工儀器網(wǎng)上看到的信息,謝謝!
穩(wěn)態(tài)模擬器及I-V測試儀,組件戶外測試系統(tǒng),紫外預(yù)處理試驗機,引出端強度試驗機,濕漏電流測試系統(tǒng),機械載荷試驗機,冰雹撞擊試驗機,旁路二極管熱性能測試系統(tǒng),熱斑耐久試驗機,EVA交聯(lián)度測試儀
UV Preconditioning Testing Chamber
Perform standards: IEC 61215:2005 & Ed.3, IEC 61646:2008, IEC 62108:2007, IEC 61345:1998
Standard | UVB of UVA+B | UVB | UVA+B |
IEC 61345:1998 | 1/2 | 7.5KW·h/m2 | 15KW·h/m2 |
IEC 61215:2005 | ≥1/3 | ≥5KW·h/m2 | 15KW·h/m2 |
IEC 61646:2008 | 3~10% | 0.45~1.5KW·h/m2 | 15KW·h/m2 |
IEC 61215 Ed.3 | 3~10% | 0.45~1.5KW·h/m2 | 15KW·h/m2 |
Note: In IEC 61646:2008 for thin-film module, need “a load sized such that at STC the module will operate near the maximum power point”, but IEC 61215 Ed.3 without this requirement. This is because if use UV-high pressure metal halide lamps or filtered xenon lamps to carry out this experiment need a load, but ultraviolet fluorescent lamps donot need. |
System integration many technical highlights, such as centralized control, automatic dimming, servo drive, full panoramic scanning, automatic monitoring irradiation uniformity, database system etc.
With touch control screen, full automatic, high precision system loop (lateral and longitudinal drive), Any mechanical action, by PLC locking. Including radiation probe Automatic full panoramic scanning, irradiance automatic integration, data acquisition and storage, automatic storage scanning data etc.
Irradiation uniformity: better than ±15%
A calibrated UV irradiation sensor (NIST trace to the source), mounted on the transmission shaft, for IEC 61215:2005 and IEC 61646:2008 to 280~385nm and 280~400nm irradiation measurement.
280~320nm to total UV ratio measured by spectrometer.
A special guide rail, can easily push and pull module.
Uniform temperature distribution, maintaining the module temperature at
Ultraviolet fluorescent lamps are cold light source without lower than 280nm UVC irradiation, to avoid the photochemical reaction of ozone synthesis, is important test conditions for IEC standard.
UVB and UVA ultraviolet fluorescent lamps, respectively for 280~385nm and 280~400nm test.
The lamps distance adjustable, asymmetric arrangement, easy to adjust the irradiation uniformity.